  • The DHCP parameter detemines if the unit uses a DHCP server to set it ethernet parameters or if it uses a user set fixed address.
  • The Fixed IP address is used if DHCP is off, or if DHCP fails.
  • In any case, the Dynamic IP address is the address being used.
  • HardyPort is the ethernet port number that one can use to connect to this instrument via UDP or TCP/IP. You send commands to the instrument to read or set parameters. The commands are text strings of the form
    PARAMETERNUMBER
    or
    PARAMETERNUMBER=PARAMETERVALUE
    For example, send the command 00006081 to read the gross weight value. The response will be something like
    Gross = 10.3
    To set a parameter, send a command like 00002887 = 0.5 - this command sets the Motion Tolerance parameter to 0.5
    See the Diagnostics/Parameters web page for a list of parameter numbers.